OMB-MULTISCAN-1200 from $2535
High Speed Isolated Temperature/Voltage Interface System

High Speed Isolated Temperature and Process Measurement System

The OMB-MULTISCAN-1200 is a compact, 19-inch rack-mountable temperature and voltage measurement interface that features channel-to-channel isolation. It has the versatility to scan thermocouples and volts at up to 147 channels/s, and digitize waveforms up to 20 kHz.

The standard unit accepts 24 differential inputs for any combination of voltage and temperature measurements. It has a 16-bit A/D converter, oversampling, and averaging for low noise and AC line rejection. Unlike most temperature measuring instruments and PC plug-in boards, which are difficult or expensive to expand beyond their built-in channel capacity, the OMB-MULTISCAN-1200 can be easily expanded up to 744 channels via low-cost slave expansion units. The OMB-MULTISCAN-1200 includes easy-to-program IEEE-488 and RS-232/422 interfaces. Multiview, a datalogging and display software and Postview, a post-acquisition display program are both provided with each unit.

SCANNING MODULES
The OMB-MULTISCAN-1200 can accept either the OMB-MTC-24 thermocouple/voltage or OMB-MHV-24 high voltage input module. Each module features 24 input channels. Using available expansion chassis, up to 30 additional input modules can be accommodated, yielding 744 total input channels. The OMB-MTC-24 thermocouple/volts module accepts J, K, T, E, R, S, B, and N thermocouple types, or ▒10 V, ▒5 V, ▒1 V, and ▒100 mV inputs. Channel-to-channel isolation is 200 Vdc peak. The OMB-MHV-24 high-voltage module accepts ▒250 V, ▒25 V, ▒2.5 V. Channel-to-channel isolation is 500 Vdc peak.

EXPANSION ARCHITECTURE
The OMB-MULTISCAN-1200 master unit can control up to 15 OMB-EXP-10A slave expansion units. Each OMB-EXP-10A accepts either on or two 24 channel input modules, providing a maximum of 744 input channels. These units may be used as bench-top units or may be mounted on a 19" rack with provided hardware.

SCANNING CAPABILITIES
The OMB-MULTISCAN-1200 provides an array of scanning capabilities to meet user application requirements. Because data logging applications frequently require the logging of readings at fixed time intervals, the OMB-MULTISCAN-1200 uses the standard hours-minutes-seconds (hh:mm:ss.s) format to specify the time interval between channel scans. Users can configure the unit to begin and end data logging on a specified event, such as a TTL signal, temperature level, IEEE GET, alarm condition, or absolute time of day, or upon completion of a specified number of readings.

ACCURACY
The unitÆs high-speed 16-bit A/D converter enables it to offer up to 0.1░C and 3.12 ╡V resolution with the OMB-MTC-24 scanning module. The OMB-MULTISCAN-1200 filters ac line cycle noise by sampling and averaging 32 measurements per line cycle, or, for more demanding applications, by averaging across multiples of 1, 2, 4, or 8 line cycles. The OMB-MTC-24 thermocouple scanning module features three strategically located temperature sensors that provide high accuracy cold-junction compensation across all 24 inputs. The OMB-MULTISCAN-1200 enables quick and accurate linearization by providing built-in lookup tables for popular thermocouple types, including J, K, T, E, R, S, B, and N. The unit can also store user-defined lookup tables in non-volatile RAM (NVRAM) for use with user-defined thermocouple linearization curves. Because individual measurements are not necessary for all data logging applications, the unit makes each channelÆs high, low, and last readings available throughout acquisition.

ALARMS
The OMB-MULTISCAN-1200 provides 32 digital alarm outputs that can be activated on a per-channel basis via user-specified alarm conditions, and it automatically returns its alarm outputs to steady state once limit conditions are resolved. The OMB-MULTISCAN-1200 can also update alarmed output channels in real-time, at its programmed scan rate, and can alert the host computer of active alarm conditions via an IEEE-488 service request (SRQ). The ability to automatically update alarm outputs in real-time affords greater control of applications and reduces programming burden by eliminating the need for constant, per-channel monitoring by the controlling computer.

MEMORY
The OMB-MULTISCAN-1200 has 256 Kbytes of data storage, which can be upgraded to 1, 4, or 8 Mbytes with standard SIMMS (single in-line memory modules). Because individual measurements are not necessary for all data logging applications, the unit makes each channelÆs high, low, and last readings available throughout acquisition.

SOFTWARE
MultiView software is provided standard with the OMB-MULTISCAN-1200. This Microsoft Windows application allows you to graphically configure channel characteristics and scanning requirements, and also provides several options for viewing, plotting, and saving measurement data. It permits the display of real-time readings in spreadsheet format and supports Dynamic Data Exchange (DDE) for transferring readings to other applications. It also enables real-time plotting of acquired readings. Postview is also included which will allow the user to recall stored data for display, and will display up to 16 channels simultaneously, and also allows the graphical data to be printed to any Windows supported printer.


animated folder

Copy (to your hard drive) Tempview and Postview software now!

View an Adobe Acrobat pdf file for more product info (requires Acrobat Reader)

View the Multiscan Quick-Start Guide (requires Acrobat Reader).

View the Multiscan complete operator's manual - 8MB (requires Acrobat Reader).